Handheld gas detection with MineARC’s Aura-PT has been designed to provide underground personnel with the ability to continuously monitor up to six gases within their immediate surroundings. When used in conjunction with the GuardIAN Intelligence Network, Aura-PT automatically communicates dangerous gas levels back to the GuardIAN Server via adjacent GuardIAN Nodes.
The Aura-PT handheld gas detector can also be used as a redundancy during re-entry, allowing personnel to easily and accurately monitor gas levels as they are driving back down the decline.
In contrast to most other DGM’s on the market, Aura-PT sensors require a service and calibration just once every 12 months.
Calibration is conducted simply by clipping the Aura-PT Calibration Shroud to the face of the unit, and connecting the gas tube. Each sensor is consequently flooded with calibration gas of known composition, and the sensor reading is adjusted to match the target accordingly.
When the sensors reach the end of their service life, they are replaced simply by removing the back of the Aura-PT, pulling out the old sensor board, and pushing in the new one. MineARC encourage a bump test following any sensor change-over to ensure the unit is operating correctly. If following a bump test, readings are out of the ordinary, calibration is advised.
Calibration can be conducted by a MineARC technician, or by an individual on-site, with the assistance of the Aura-PT manual and calibration kit.
Charging
The Aura-PT Handheld Gas Monitor is MineARC’s answer to localised gas detection and alerts for personnel. The MineARC Charger Rack can charge up to 50 Aura-PT Handheld Gas Monitors simultaneously on one side.
Monitors per row: 10
Single sided max: 50
Double sided max: 100
FAQs
Q: WHERE IS AURA-PT MANUFACTURED?
Aura-PT is designed and constructed by MineARC Systems in Perth, Western Australia. The system utilises quality components from companies around the world.
Q: WHAT IS THE DIFFERENCE BETWEEN A BUMP TEST AND CALIBRATION?
The main difference is that a bump test determines whether the Aura-PT can detect if a possibly hazardous gas is present, while calibration checks that equipment is accurate.
Bump Test: Checks whether sensors and alarms are working as intended, and failure might indicate that a blockage is present. In sum, bump testing assesses function, not accuracy.
Calibration Check: Calibration checks start by “zeroing” the Aura-PT and testing that alarms go off after applying a high enough concentration of test gas. The resulting sensor reading should match the concentration listed on the test gas container. The device is accurate within an acceptable range of ±10-20% of the test gas concentration applied.
Full Calibration: The adjustment of the sensor’s response to match the desired value compared to a known traceable concentration of test gas. This adjustment accounts for naturally occurring drifting and other environmental factors. Specially trained, qualified personnel are the only people permitted to perform full calibrations.
Q: WHAT IS THE DIFFERENCE BETWEEN A BUMP TEST AND CALIBRATION?
The sensor life remains the same, regardless of if the unit is left on or off.
Digital Gas Monitoring Node – GuardIAN Gas Node Digital Gas Monitoring Nodes from the GuardIAN range provide a remote gas monitoring solution for mine sites; designed specifically to be mounted to rock wall and high traffic areas of the underground mine. Sensor technology within each node monitors gas levels in surrounding areas; designed to facilitate
Underground Tracking Nodes – GuardIAN Designed specifically for use in underground environments, MineARC’s Personnel Tracking Node system ensures all workers safety and positioning is monitored at all times. Underground Tracking Nodes from the GuardIAN range has been specifically developed to integrate with the GuardIAN Intelligence Network, allowing sites to remotely monitor the location and well-being
Aura-FX Digital Gas Monitoring Aura-FX Digital Gas monitoring is an essential aspect of maintaining a life-supporting atmosphere within the refuge chamber. During entrapment, occupants consume oxygen and expire carbon dioxide (CO2) and carbon monoxide (CO) as part of their normal respiration. CO can also enter the chamber during occupant entry and if the compressed air
We use cookies to ensure that we give you the best experience on our website. If you continue to use this site we will assume that you are happy with it.AcceptDenyPrivacy policy